0

Out of 0 Ratings

Owner's of the Leica Vacuum Cleaner ASP300 gave it a score of 0 out of 5. Here's how the scores stacked up:
  • Reliability

    0 out of 5
  • Durability

    0 out of 5
  • Maintenance

    0 out of 5
  • Performance

    0 out of 5
  • Ease of Use

    0 out of 5
of 70
 
58
Instruction manual V 1.3 – 12/2003
6.4 Typical fill or drain problems
Fill and drain problems may be due to several causes:
1. There is insufficient reagent
Verify that both the reagent bottles and the wax stations have
sufficient reagent.
2. Insufficient pressure or vacuum
The Leica ASP300 is unable to create sufficient vacuum (for a fill) or
pressure (for a drain) in the retort.
Verify that the reagent bottle in question is correctly inserted in
the reagent cabinet.
Verify that all reagent bottles are fully pushed home into their
sockets.
Check whether there is some debris under the retort lid seal
causing a leak.
3. There is a blockage in the air or reagent lines
If there is reagent in the retort that cannot be drained, try to remove
the blockage as follows:
Heat the retort to the maximum possible temperature (consistent
with the reagent currently in the retort, if any) and leave it at this
temperature for at least 15 minutes.
Whilst the retort is at maximum temperature, attempt to fill or
drain.
If the retort is empty:
Run the extended retort clean program. Use the maximum set-
tings for number of clean cycles and temperature (consistent
with the type of reagents/solvents used in the clean program).
If you cannot eliminate the problem by carrying out the above
listed steps, call Leica Technical Service
6. Trouble shooting

   Find Your Products By Category