Leica ASP300
Program Options
Selection of the options on how programs are carried out.
STATION ALLOCATION: By age or Sequential
By age - RMS is activated, reagents will automatically be
used in order of increasing cleanliness.
Sequential - reagents will be used in sequential order of stations.
WAX BATH ORDER: Auto or 1.; 2. ; 3.
Wax bath order can be activated only when STATION ALLOCATION
is set to Sequential.
PROMPT FOR NUM BLOCKS : Enabled or Disabled
Enabled- When starting a program, the actual number of
blocks must be entered.
This is recommended with the RMS being enabled.
Disabled - Number of blocks can be entered optionally (not
mandatory).
Date / Time
Be sure to verify that the date and time entered do actually correspond
to the local time/date.
Press the buttons to access the entry windows where time/date can be
corrected, if necessary.
Instrument
Press WAX BATH SET TEMP (°C) to access the wax bath temperature
entry window. Select the highest temperature compatible with the type
of paraffin used as paraffin temperature will drop slightly once the paraf-
fin is in the retort.
5. Operation