Instruction Manual V 1.3 – 12/2003
Fig. 14
Remove the splash protector.
Fill the wax bath (Fig. 15) with wax pellets up
to the top.
Press BACK and REAGENT STATUS to ac-
cess the reagent status table (Fig. 14). High-
light the wax bath you just filled and press
SET AS FULL.
After about 90 minutes, add another 0.5 kg
of wax pellets.
The pellets will take a total of approx. 10
hours to melt.
Important!
During the melting process, the paraf-
fin level MAY NOT drop below the
MIN level (2, Fig. 15).
Once the paraffin is completely molten, re-
insert the splash protector.
Software control
The software calculates when the wax will
be ready for use and selects program start-
ing times accordingly, i.e. the paraffin will
definitely be molten when it is required for
the first paraffin step.
Filling with wax pellets