6.1 General
The Leica ASP300 is equipped with a trouble shooting system.
If an error is detected, detailed error messages and corresponding user
instructions on what to do are displayed on the touch screen.
For all correctable error conditions the instrument will take
corrective actions to protect the specimens.
Details of all events are recorded in “Run Log”.
Run log can be viewed by pressing MENU and then RUN LOG.
6.2 Power failure
Power failure during a processing program
In the event of a power failure, the instrument automatically compen-
sates any impact the power failure might have on the specimens.
Power failure during a paraffin or retort clean cycle
If a power failure occurs during a clean cycle, the total duration of
the cycle will be extended by the duration of the power failure.