Instruction manual V 1.3 – 12/2003
Allocation of reagents to the correct reagent group is the basis for
compatibility monitoring. Allocation to the wrong reagent group
can lead to reagent cross contamination.
Entering / modifying reagent thresholds
Highlight the reagent to be modified.
Press the headline key symbol of the parameter to be changed.
Enter a new threshold or - if no warning is desired - press CLEAR to
remove the threshold altogether.
Press OK to confirm.
The same threshold value applies for all reagent stations con-
taining the same reagent.
Changing reagent names or reagent groups
Highlight the reagent.
Press the corresponding headline key symbol.
In the entry window (or via the keyboard), enter the new reagent
group allocation.
Press OK to confirm and save.
Reedit all station and programs allocated to the reagent changed.